MCO-18AC
Ambient Temperature : 5°C to 50°C
Effective Capacity : 170 liters (6.0 cu.ft.)
| Model | MCO-170AIC |
|---|---|
| External Dimensions | 620 x 710 x 900 (mm) / 24.4 x 28.0 x 35.4 (inch) |
| Interior Dimensions | 490 x 523 x 665 (mm) /19.3 x 20.6 x 26.2 (inch) |
| Net Weight | 80kg |
| Interior Volume | 165 liters (5.8 cu.Ft.) |
| Data Acquisition | Automatic log function of temperature, CO2, Door opening/closing, Alarm and CSV file output |
| Exterior Cabinet and Door | Galvanized steel with backed-on finish |
| Interior and Shelves | Copper-enriched stainless steel |
| Outer Door | Galvanized steel with backed-on finish, Reversible heated |
| Inner Door | Tempered glass |
| Shelves | Copper-enriched stainless steel, 4 supplies as standard (Maximum 10), Exterior dimensions: 475 (W) x 450 (D) x 12 (H) mm, maximum load 7kg/shelf, Adjustable |
| Access Port | Diameter 30mm port with non-VOC silicone stoppers (1 on back side) |
| Insulation | Rigid foam polyurethane |
| Air Flow | Gentle vertical airflow, continuous with inner door closed |
| Air Filter | 0.3μm, Effeciency:99.97% or higher (for CO2) |
| Communication | Remote alarm contacts standard. Optional 4-20mA connection. Optional with RS-232C/ RS-485/LAN data ports |
| CO2 Gas Connection | 4mm to 6mm inner diameter tubing |
| CO2 Gas Pressure | 0.03 MPa (G) (0.3Kgf/cm2G, 4.3psiG) from two stage CO2 regulator |
| Tempearture Control Range | +5°C above ambient to 50 °C*1 (Ambient temperature: 5°C-35°C) |
| Tempearture Uniformity | ±0.25°C (23°C ambient, setting: 37°C, CO2: 5%, no load)*2 |
| CO2 Control Range and Deviation | 0% to 20%,±0.15% (23°C ambient, setting 37°C, 5% CO2, no load) |
| Interior Humidity | 95% ±5%R.H. at 37°C by natural evaporation with humidifying pan |
| Max. Power Consumption | Max. 379W |
| Max. Heat Discharge | 1,070kJ/h |
| CO2 Sensor Platform | Ceramic based, single beam infrared sensor, with dual wavelength measurement for continuous auto-zero calibration |
| CO2 Sampling Patent Pending | No moving parts; airflow passess over in/out ports to sustain continuous sampling |
| Leveling Feet | 4 |